Development and Performance Study of a Capillary Liquid Scintillator Neutron Detector

请选择邀稿期刊:
摘要: Capillary liquid scintillator detectors are promising for high-resolution neutron imaging, yet experimental data on their light spread mechanism and spatial performance remain limited. Here, we report a neutron detector based on a hexagonal capillary array filled with EJ-309 liquid scintillator, with an inner diameter of about 50 um and a camera readout of 9 um pixels. Laser experiments show that the FWHM of the full light spot decreases from 260 um to 90 um with a metal light absorber, confirming effective suppression of lateral light spread. Using an AmBe neutron source, an effective field of view with a 5-sigma threshold was established from background frames. For single-capillary events, the pulse height spectrum follows a Landau distribution with a most probable value of 0.133 +/- 0.001 (stat.), and the intrinsic detection efficiency is 10.07% +/- 1.26% (stat.) +/- 1.43% (syst.), corresponding to about 13.55% when normalized to the active liquid scintillator area. The point spread function core yields a radial FWHM of 12.8 um and a centroid positioning precision of approximately 5.5 um (1 sigma), while the intrinsic position resolution is limited by the capillary pitch to 54 um. Linearity is good for 1 to 2 capillaries, with deviation appearing for 2 to 3 capillaries due to additional capture of spread light. These results provide experimental basis and physical understanding for imaging applications of capillary liquid scintillator neutron detectors.