您当前的位置: > 详细浏览

Designing a multi-reflection time-of-flight mass analyzer for LPT

请选择邀稿期刊:
摘要: A new method including two sub-procedures, global search and local refine, has been developed and presented to design a multiple-reflection time-of-flight (MRTOF) mass analyzer. By using this method, a different type of MRTOF mass analyzer, in which each mirror consists of five cylindrical electrodes, has been designed for isobaric separation for Lanzhou Penning Trap (LPT). The optimal potential parameters of the electrodes have been obtained and the maximal resolving power has been achieved to be 1.3x105 with a total time-of-flight of 6.5 ms for an ion species of 40Ar1+. The simulation shows the inaccuracy of the potentials applied to the mirror electrodes must be less than 50 ppm or preferably 20 ppm.

版本历史

[V1] 2023-06-24 09:09:11 ChinaXiv:202306.00650V1 下载全文
点击下载全文
预览
同行评议状态
通过
许可声明
metrics指标
  •  点击量2228
  •  下载量235
评论
分享