您当前的位置: > 详细浏览

Variance analysis for passive neutron multiplicity counting

请选择邀稿期刊:
摘要: Passive neutron multiplicity counting is widely used as a nondestructive assay technique to quantify mass of plutonium material. One goal of this technique is to achieve good precision in a short measurement time. In this paper, we describe a procedure to derive mass assay variance for multiplicity counting based on the three-parameter model, and analytical equations are established using the measured neutron multiplicity distribution. Monte Carlo simulations are performed to evaluate precision versus plutonium mass under a fixed measurement time with the equations. Experimental data of seven weapons-grade plutonium samples are presented to test the expected performance. This variance analysis has been used for the counter design and optimal gate-width setting at Institute of Nuclear Physics and Chemistry.

版本历史

[V1] 2023-06-18 18:11:07 ChinaXiv:202306.00313V1 下载全文
点击下载全文
预览
许可声明
metrics指标
  •  点击量502
  •  下载量152
评论