您当前的位置: > 详细浏览

Segmented TOF at 7 T MRI: Technique and clinical applications 后印本

请选择邀稿期刊:
摘要: Purpose: Time-of-flight (TOF) MR angiography has an advantage of contrast and resolution in ultra-high field (7 T) MRI systems. However, increased specific absorption rate (SAR) prohibits the application of spatial saturation band, leading to venous contamination in maximum intensity projection (MIP) images. Methods: A segmented k-space filling scheme with sparse venous saturation pulses was developed for 7 T TOF-MRA. The effectiveness of the segmented TOF sequence was verified by Bloch equation simulation and experiments on 3 T. The protocol on 7 T was optimized and applied for healthy volunteers and patients with vascular diseases. Results: Segmented TOF achieved equivalent contrast and venous suppression effect as conventional methods, while SAR values had a remarkable reduction and obeyed the limit of a 7 T MRI system. The decreased number of saturation pulses allowed shorter acquisition time than existing solutions. The comparison of segmented TOF and conventional TOF revealed flow direction in vascular diseases. Conclusion: Segmented TOF is proved to be a time-efficient way to achieve high-resolution angiograms without venous contamination at ultra-high field. The sequence holds strong promise for non-contrast clinical diagnosis on cerebrovascular diseases. (C) 2015 Elsevier Inc. All rights reserved.

版本历史

[V1] 2016-05-12 08:40:39 ChinaXiv:201605.01416V1 下载全文
点击下载全文
预览
同行评议状态
待评议
许可声明
metrics指标
  •  点击量2680
  •  下载量1378
评论
分享