• An improved method for analyzing the output stability of medical LINAC based on planar dose

    分类: 核科学技术 >> 粒子加速器 提交时间: 2023-06-18 合作期刊: 《Nuclear Science and Techniques》

    摘要: The absolute dose difference of the iso-center position at different times was used to evaluate the stability of medical linear accelerator (LINAC). In order to eliminate the drawback of point dose without statistical significance and to investigate the output stability of segment with small weight related to intensity-modulated radiotherapy (IMRT), a modified method of evaluating the stability of LINAC was proposed based on planar dose combined with gamma method. With a commercial available two-dimensional ionization chamber, a set of planar doses with dose gradient from 2 cGy to 100 cGy were obtained. Then the gamma method was adopted to analyze the dose difference between the reference data and the evaluated data at each dose level. The results showed that the improved method based on planar dose for analyzing the output stability of the medical LINAC was feasible and efficient, and suggested that the reverse optimization should be aborted in clinical when the segment weight related to IMRT was under 10 MU.

  • Radiotherapy reliability analysis based on PSA method

    分类: 物理学 >> 核物理学 提交时间: 2023-06-18 合作期刊: 《Nuclear Science and Techniques》

    摘要: The reliability of radiotherapy was evaluated and effective approaches were obtained in order to improve radiotherapy quality by using the Probabilistic Safety Assessment (PSA) method. This study investigated the feasibility of the PSA method being applied to radiotherapy through Image-guided Radiotherapy (IGRT) and chest tumor irradiation. A fault tree has been constructed after analyzing causal relationship of the events. After calculating RiskA, a total inaccuracy radiotherapy probability and the importance of all base events were obtained. The probability of inaccurate radiotherapy was 2.87%. Under the condition that the target delineation was perfectly right, the accuracy of radiotherapy significantly improved. With the calculation without Cone-beam Computed Tomography (CBCT) being corrected before irradiation, the accuracy significantly decreased. The most important events were connected with the human factor. Improving human technical level could enhance radiotherapy quality control efficiently.